Description
The new improved compact design includes an automatic detection of the initial boiling point (IBP) as well as quick connections for parts to be removed after operation.
The distillation is performed from the Initial Boiling Point (IBP) to the End Boiling Point (EBP) by the operator. The criteria for a distillation end are:
- The EBP in AET-temperature is exceeded
- The distillate volume is exceeded
- The maximum limits of heating bath temperature or flask temperature are exceeded
- Product cracking in the flask
- Vacuum loss
The distillate volume has to be measured by the operator in a temperature controlled receiver.
A safety enclosure of the system to adequately shield the operator from the distillation apparatus in case of mishap as suggested by ASTM D1160-15 is quoted as an option as some customers will use the system within a fumehood.
Technical Data
Available flask sizes | 500 mL |
Flask charge | 200 mL |
Max. flask temperature | 400° C (750° F) |
Max ambient temperature | 25° C |
Operating pressure | vacuum down to 1 Torr |
Mains supply | 208 – 250 V, 50 Hz (60 Hz upon request) |
Power consumption | 3500 W (without options) |
Dimensions (w x h x d) | approx 0.65 x 0.98 x 0. m |